24.0.3 One (1) gravity type Thorpe tube oxygen flow meter shall be provided. This <br />flow meter shall be certified by the equipment manufacturer to meet the appropriate <br />FDA standards for medical oxygen flow meters. <br />Y,/N� <br />24,1 REGULATOR <br />A 50 psi, preset medical oxygen regulator shall be provided and installed on the <br />supply end of the oxygen system hose. This regulator shall be certified by the <br />equipment manufacturer to meet the appropriate FDA standards for medical oxygen <br />flow meters. <br />Y_j/N <br />24.2 OXYGEN CONTENT GAUGE <br />A digital oxygen content gauge with an alarm shall be provided and wired to the <br />rotary timer. <br />Y-A.A_ <br />24.3 CARBON MONOXIDE MONITOR <br />A carbon monoxide monitor with an alarm shall be provided and installed in the <br />action area, within the patient compartment. <br />Y-ZV <br />24.4 ELECTRIC OXYGEN SHUT OFF <br />An electric shut off switch for the oxygen system shall be provided and installed in <br />the rear action area switch panel. The system shall include a manual bypass. <br />YjZN_ <br />24.5 PORTABLE/ ONBOARD SUCTION AND ASPIRATION SYSTEM <br />A portable/onboard, complete self-contained S-SCORT VX-2 dual mode unit with <br />battery charger shall be installed on the primary action area wall. It shall include a <br />charging retention bracket hard -wired to 12 VDC power. A 1,200 cc suction canister, <br />6' suction tubing, and suction tip shall be supplied with unit. <br />Y._iG.N <br />24.6 SHARPS/E310-HAZARD <br />A sharps container with a crash stable mounting bracket shall be supplied and <br />shipped loose. A biohazard container with lid and twelve [12] red biohazard bags <br />shall be shipped loose for installation by the department. <br />Y/O N <br />24.7 PATIENT CODE SIGNAL SYSTEM <br />In the action area switch console there shall be three [3] colored switches (red, <br />yellow and green) that activate corresponding indicator lights in the driver's switch <br />console. In addition, there shall be a switch in the action area switch console that <br />activates a buzzer in the driver's compartment. These lights and buzzer shall be <br />used to notify the driver of any change in the patient's status. <br />Y <br />24.8 RADIO WIRING <br />57 <br />