<br />The LIFEPAK 15 monitor/defibrillator has six main
<br />operating modes:
<br />AED Mode: for automated ECG ana$ysis and a prompted
<br />treatment protocol for patients in cardiac arrest.
<br />Manual Mode: for performing manual defibrillation,
<br />synchronized cardioversion, noninvasive pacing, and
<br />ECG and vital sign monitoring.
<br />Archive Mode: for accessing stored patient information.
<br />Setup Mode: for changing default settings of the
<br />operating functions.
<br />Service Mode: for authorized personnel to perform
<br />diagnostic tests and calibrations.
<br />Demo Mode: for simulated waveforms and trend graphs
<br />for demonstration purposes.
<br />Weight:
<br />Basic monitor/dolibrillalor with now roll paper and two
<br />batteries installed: 7.9 kg (17,5 lot
<br />Fully featured monitorldefibrillator with new roll paper
<br />and two batteries installed: 8.4 kg (18.5lb)
<br />Lithium -Ion battery: sO.Bkg (1.31'b)
<br />Accessary Bags and Shoulder Strap:1.77 kg (3.9 lb)
<br />Standard (hard) Paddles: 0,95 kg 12.1 lb)
<br />Height: 31.7 cm (12.5 in)
<br />Width: 40.1 cm (15.8 in)
<br />Depth: 23.1 cm (9.1 in)
<br />Size (active viewing area): 212 mm (8.4 in) diagonal;
<br />171 mm (6.7 in) wide x 1128 mm (5.0 in) high
<br />Resolution: display type 640 dot x 480 dot color backlit LCD
<br />User Seiectable Display Mode; full color or SunVueTn"
<br />display high contrast
<br />Display: a minimum of 5 seconds of ECG and alphanumerics
<br />for values, device instructions, or prompts
<br />Display: up to three waveforms
<br />Waveform Display Sweep Speed: 25 mm/sec for ECG,
<br />SpO,, IP, and 12.5 rem/sco for CO,
<br />[IM11MIZE,
<br />MINUMINEEM
<br />The device captures and stores patient data, events (including
<br />waveforms and annotations), and continuous waveform and
<br />patient impedance records in internal memory.
<br />The user can select and print reports, and transfer the stored
<br />information via supported communication methods.
<br />Report Types;
<br />- Three format types of CODE SUMMARYTr^ critical event
<br />record: short, medium, and long
<br />- 12-lead ECG with STEMI'statements
<br />Continuous Waveform (transfer only)
<br />-- Trond Summary
<br />- Vital Sign Summary
<br />- Snapshot
<br />Memory Capacity: Total capacity is 360 minutes 0f
<br />continuous ECG, 90 minutes of continuous data from all
<br />channels, or 400 single waveform events..
<br />Maximum memory capacity for a single patient includes up to
<br />200 single waveform reports and 90 rninules 0f continuous ECG.
<br />The device is capable of transferring data records by wired
<br />or wireless connection. This device compiles with Part 15
<br />of the FCC rules„ and its operation is subject to the following two
<br />conditions: (1) this device may not cause harmful interference,
<br />and (2) this device must accept any interference received,
<br />including interference that may cause undesired operation.
<br />Serial Port RS232 communication + 12V available
<br />Limited to devices drawing maximum 0.5 A current
<br />Bruotooth° technology provides short-range wireless
<br />communication with other Bfuetooth-enabled devices
<br />EGG
<br />ECG is monitored via several cable arrangements:
<br />A 3-wire cable is used for 3-lead ECG monitoring.
<br />A 5-wire cable is used for 7-lead ECG monitoring.
<br />A 1 O-wire cable is used for 12-lead ECG acquisition. When
<br />the chest electrodes are removed, the 110-wire cable functions
<br />as a 4-wire cable.
<br />Standard paddles or QUIK-COMBO pacing/defibrillation/ECG
<br />electrodes are used for paddles lead monitoring.
<br />Frequency Response:
<br />Monitor: 0.5 to 40 Hz or 1 to 30 Hz
<br />Paddles: 2.5 to 30 Hz
<br />12-lead ECG diagnostic: 0.05 to 150 Hz
<br />Lead Selection:
<br />Leads I, II, III, (3-wire ECG cable)
<br />Leads I, II, Ili, AVR, AVL, and AVE acquired simultaneously
<br />(4-wire ECG cable)
<br />Lead's 1, II, Ill, AVR, AVL, AVF„ and C lead acquired
<br />simultaneously (5-wire EGG cable)
<br />Leads 1, II, III, AVR, AVL, AVF, V1,V2,V3,V4,V5, and V6
<br />acquired simultaneously it0-wireECG cable)
<br />ECG size: 4, 3, 2.5,, 2,1.5,1, 0.5, 0.25 cm/mV
<br />(fixed at ii cm1mV for 12-lead)
<br />Heart Rate Display:
<br />20-300 bpm digital display
<br />Accuracy: ±4% or ±3 bpm, whichever is greater
<br />QRS Detection Hangs Duration: 40 to 120 msec
<br />Amplitude: 0.5 to 5.0 m
<br />Common Made Rejection (CMRR)a ECG Leads: 90 dB
<br />at 50160'Hz.
<br />spol'1SPGOPSIIism t
<br />Sensors
<br />MASIMO sensors including RAINBOW" sensors
<br />NELLCOR" sensors when used with the MASIMO REDTM
<br />MNC adapter
<br />SII1102
<br />Displayed Saturation Range: "<50" for levels below
<br />50%;50to100%
<br />Saturation Accuracy: 70-100%(0-69%uunspecified)
<br />Adults/Pediatrics:
<br />±2 digits (during no motion conditions)
<br />,3 digits (during motion conditions)
<br />Dynamic signal strength bar graph
<br />Pulse tone as Sp02 pulsations are detected
<br />SpO Update Averaging Rate User selectable:
<br />4, 8,12 or 16 seconds
<br />Sp02 Sensitivity User selectable: Normal, High
<br />SpOZ Measurement: Functional Sp02 values are
<br />displayed and stored
<br />Pulse Rate Range:25 to 240 bpm
<br />Pulse Rate Accuracy (Adults/Pediatrics):
<br />±3 digits (during no:noUon conditions)
<br />±5 digits (during motion conditions)
<br />Optional Sp02 waveform display with autogain control
<br />tiptAi '
<br />SpCO Concentration Display Range: 0 to 40%
<br />SpCO Accuracy; ±3 digits
<br />SPME;f°"
<br />SpMet Saturation Range:0 to 15.0%
<br />SpMet Display Resolution:0.1%a up to 10%
<br />SpMet Accuracy: ±1 digit
<br />MBP
<br />Blood Pressure Systolic Pressure Rangm 30 to 255 mmHg
<br />Diastolic Pressure Range:16 to 220 mmHg
<br />Mean Arterial Pressure Range: 20 to 235 mmHg
<br />Units: mmHg
<br />Blood Pressure Accuracy: ±5 mmHg
<br />Blood Pressure Measurement Time: 20 seconds, typical
<br />(excluding cuff inflation time)
<br />Pulse Rate Range: 30 to 240 pulses per minute
<br />Poise Rate Accuracy: ±2 pulses per minute or t2%,
<br />whichever is greater
<br />Operation Features Initial Cuff Pressure: User selectable,
<br />80 to 180 mmHg
<br />Automatic Measurement Time Interval: User selectable,
<br />from 2 min to 60 min
<br />Automatic Cuff Deflation Excessive Pressure: If cuff
<br />pressure exceeds 290 mmHg
<br />Excessive Time: If measurement time exceeds 120 seconds
<br />co,
<br />CO2 Range: 0 to 99 mmHg (0 to 13.2 kPa)
<br />Units: mmHg,, %, or kPa
<br />Respiration Rate Accuracy:
<br />0 to 70 bpm: ±-4 bpm
<br />71to99bpm: ±2bpm
<br />Respiration Rate Range: 0 to 99 breaths/minute
<br />Rise Time;190 msec
<br />Response Time: 3.3 seconds (Includes delay time
<br />and rise time)
<br />Initialization Time: 30 seconds (typical), 10-180 seconds
<br />Ambient Pressure: automatically compensated internally
<br />Optional Display: CO2 pressure waveform
<br />Scale factors: Autoscale, 0-20 mmHg (0-4 Vo$%a),
<br />0-50 mmHg (0-7 V016/6), 0-100 mmHg! (0-14 Vol%o)
<br />invuashm, Piie,ssuure
<br />Transducer Type: Strain -gauge resistive bridge
<br />Transducer Sensitivity: 5pV/VlmmHg
<br />Excitation Voltage: 5 Vdc
<br />Connector: Efectro Shield: CXS 3102A 14S-6S
<br />Bandwidth: Digital tittered, DC to 30 Hz Lc -3db)
<br />Zero Drift: 1 mmHg/hrwithout transducer drift
<br />Zero Adjustment: -+150 mmHg including transducer offset
<br />Numeric Accuracy: ±1 mmHg or 2% of reading, whichever
<br />Is greater, plus transducer error
<br />Pressure Range: -30 to 300 mmHg, in six user selectable ranges
<br />h'avasive Pressure Display
<br />Display: IP waveform and numerics
<br />Units: mmHg
<br />Labels: P'1 or P2, ART, PA, CVP, ICP, LAP (user selectable)
<br />"Temperature
<br />Range: 24.8° t0 45.2°C (76.6° to 113.4'F)
<br />Resolution: 0.1"C
<br />Accuracy: ±0.2°C including sensor
<br />Reusable Temperature Cable: 5 foot or 10 foot
<br />Disposable Sensor Types; Surface -Skin;
<br />Esophageal/Rectal